Inomax Cylinder D-Size Duration Chart - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system, series 3 software
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

Duration Chart
(D-size)
INOMAX Cylinder D-Size
5
10
20
40
80
This chart is representative of a range of doses available
on the INOmax DS
intended as the recommended therapeutic dose.
* All calculations in the table above are based on a full cylinder, 138 bar (2000
psig), 353 liter "D" cylinder, with a cylinder change at 14 bar (200 psig). The
figures are calculated based on a total continuous breathing circuit gas flow
and a cylinder conversion factor of 2.6 liters per bar/0.18 liters per psig.
• INOMAX flow = [Desired dose × total ventilator flow] ÷ [Cylinder concentration
- desired dose]
• Cylinder volume = Cylinder conversion factor × cylinder pressure (bar/psig)
• Cylinder duration (hours) = (Cylinder volume ÷ INOMAX flow rate) ÷ 60
Calculations are considered estimates and may vary under clinical
circumstances. For more information, call 1-877-KNOW-INO (1-877-566-9466).
Part No. 20751 Rev-01
2014-08
5 L/min
10 L/min
7.0 Days
3.5 Days
3.5 Days
1.7 Days
1.7 Days
20.7 Hours 10.3 Hours 5.2 Hours
20 Hours
10 Hours
9.5 Hours
4.8 Hours
Typically used in transport
and doses higher than 20 ppm are not
IR
16
FLOW
20 L/min
1.8 Days
21 Hours 10.5 Hours
5 Hours
2.4 Hours
40 L/min
21 Hours
2.5 Hours
1.2 Hours