Connection To Various Breathing Systems - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system, series 3 software
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

Proper use of these products depends on careful reading and
understanding of labeling and instructions. Please refer to the
INOmax DS
and INOblender operation manuals for guidance. Also
IR
refer to the specific breathing device operation manual or instructions
for use.
INOmax DS
• INOmax DS
IR
gas sampling system at 230 mL per minute; this can affect the
sensitivity of a flow triggered synchronized breath mode of
some ventilators. The trigger sensitivity of the ventilator should
be checked after connecting the INOmax DS
circuit.
• Patient disconnect and high-pressure alarms are required for
the ventilator.
INOmax DS
• Insert the Injector Module on the dry side of the breathing circuit
prior to the humidifier (this will ensure correct flow measurement).
• Avoid medications interfering with the gas monitoring system;
administer any aerosolized medications distal to the sampling tee.
Warnings:
IR
subtracts gas from the breathing circuit via the
Cautions:
IR
Circuit Connection
to the breathing
IR
Part No. 20751 Rev-01