Changing INOMAX
Cylinders

Changing INOMAX Cylinders

WARNING:
• A new INOMAX cylinder and regulator must be purged
before use to ensure the patient does not receive an
excess level of NO
• Loss of communication between the INOmax DS
the INOMAX cylinder for more than one hour will result in
interruption of INOMAX delivery.
Caution:
• Replace an INOMAX cylinder when its pressure is
less than 200 psig.
• When using the Transport Regulator/Cap Assembly (PN 10022)
ensure the cap is fully seated and in place on the INOmeter
and the infrared cable is connected and latched to the infrared
connector port on the back of the INOmax DS
Note: Ensure the
white plastic tip
is in place.
A. Attach a regulator
to an INOMAX
cylinder with
greater than
500 psig.
Part No. 20751 Rev-01
2014-08