Probe Compensation; Probe Compensation - Tektronix 2215A Instruction Manual

Hide thumbs Also See for 2215A:
Table of Contents

Advertisement

Operating Instructions— 221
should not be required. If adjustment is needed, perform the
following procedure:
1. Preset instrument controls and obtain a baseline trace.
2.
Use the Channel 1 POSITION control to move the
baseline trace to the center horizontal graticule line.
3.
If the resulting trace is not parallel to the center hori­
zontal graticule line, use small flat-bit screwdriver to adjust
the TRACE ROTATION control and align the trace with the
center horizontal graticule line.

PROBE COMPENSATION

Misadjustment of probe compensation is a common
source of measurement error. Most attenuator probes are
equipped with a compensation adjustment. To ensure opti­
mum measurement accuracy, always compensate the oscil­
loscope probes before making measurements. Probe
compensation is accomplished as follows:
1. Preset instrument controls and obtain a baseline trace.
2.
Connect the two 10X probes (supplied with the instru­
ment) to the CH 1 and CH 2 input connectors.
3.
Set both VOLTS/DIV switches to 10 mV and set both
Input Coupling switches to DC.
4.
Select CH 1 VERTICAL MODE and insert the tip of the
Channel 1 probe into the PROBE ADJUST output jack.
5.
Using the approximately 1-kHz PROBE ADJUST
square-wave signal as the input, obtain a 5-division display
of the signal.
2-10
Service
5A
6.
Set the A SEC/DIV switch to display several cycles of
the PROBE ADJUST signal. Use the Channel 1 POSITION
control to vertically center the display.
7.
Check the waveform presentation for overshoot and
rolloff (see Figure 2-9). If necessary, adjust the probe com­
pensation for flat tops on the waveforms. Refer to the in­
structions
supplied
with
the
compensation adjustment.
8.
Select CH 2 VERTICAL MODE and connect the Chan­
nel 2 probe tip to the PROBE ADJUST output jack.
9.
Use the Channel 2 POSITION to vertically center the
display and repeat step 7 for the Channel 2 probe.
10. Disconnect the probes from the instrument.
KX
9<
1C
....
Figure 2-9. Probe compensation.
probe
for
details
of
CORRECT
(FLAT)
OVER C O M ­
PENSATED
(OVERSHOOT)
r ~
UNDER C O M ­
PENSATED
(ROLLOFF)
C4732-15

Advertisement

Table of Contents
loading

Table of Contents