Horizontal - Tektronix 2215A Instruction Manual

Hide thumbs Also See for 2215A:
Table of Contents

Advertisement

Equipment Required (see Table 4-1):
Calibration Generator (Item 1)
Leveled Sine-Wave Generator (Item 2)
Time-Mark Generator (Item 3)
INITIAL CONTROL SETTINGS
Vertical
POSITION (both)
VERTICAL MODE
BW LIMIT
CH 1 VOLTS/DIV
CH 1 VOLTS/DIV
Variable
Channel 1 Input Coupling

Horizontal

POSITION
HORIZONTAL MODE
A SEC/DIV
SEC/DIV Variable
X I0 Magnifier
B DELAY TIME POSITION
B TRIGGER
SLOPE
LEVEL
A TRIGGER
VAR HOLDOFF
Mode
SLOPE
HORIZONTAL
Midrange
CH 1
Off (button out)
0.5 V
CAL detent
DC
Midrange
A
0.05 #xs
CAL detent
Off (knob in)
Fully counterclockwise
OUT
Fully clockwise
NORM
NORM
OUT
Performance Check Procedure— 2215A Service
50-0 BNC Cable (Item 4)
50-0 BNC Termination (Item 5)
LEVEL
A&B INT
A SOURCE
A EXT COUPLING
PROCEDURE STEPS
1. Check Timing Accuracy and Linearity
a.
Connect the time-mark generator output via a 50-0
cable and a 50-0 termination to the CH 1 OR X input
connector.
b.
Select 50-ns time markers from the time-marker
generator.
c.
Adjust the A TRIGGER LEVEL control for a stable,
triggered display.
d.
Use the Horizontal POSITION control to align the sec­
ond time marker with the second vertical graticule line.
e.
CHECK— Timing accuracy is within 3% (0.24 division
at the 10th vertical graticule line), and linearity is within 7%
(0.14 division over any 2 of the center 8 divisions).
Midrange
VERT MODE
INT
DC -- 10
4-7

Advertisement

Table of Contents
loading