13.3 Exiting Injection Complete

13.3.1 Conducting Another Injection

NOTE: Depending on how Fluid Delivery Setup is configured, the protocol may reset after the injection. The
default setting is for the system to keep the previous protocol for the next injection.
1.
Select Same Patient.
a.
The operator is returned to the Home screen.
b. See
"Section 10.1 - Create or Edit a Protocol"
Arming the
Injector".
2.
Select New Patient.
a.
The following message will display. Disconnect the patient, remove syringes from the injector, then press OK.
b. The operator is returned to the Home screen.
c.
See
"Chapter 11 - Preparing for Injection"
to create or edit a protocol, or repeat the steps in
for further instructions.
13 - 57
"Section 12.2 -