10 - 38
3.
Select two of the three parameters (3)
information) and enter the values using the keypad(4). The third parameter will be calculated automatically.
4.
Select Enter to confirm the entered value or select Cancel to disregard.
5.
Repeat step 1 to add additional phases.
NOTE: The system provides an on-screen indication and displays how much fluid needs to be added in each
syringe if the programmed volume exceeds the amount of volume in the syringe(s). Alternatively, the
recalculated values can be accepted on the Insufficient Volume Screen.
NOTE: When modifications have been made to a protocol, an asterisk (*) appears to the right of the protocol
name.
NOTE: The protocol can be saved (see
"Chapter 12 - Arming and
6.
Optionally, modify the default pressure limit.
a.
Select Pressure Limit (1) from the Home Screen.
b. Select a pressure limit from the pick list(2).
NOTE: Ensure that the pressure is set for the patient per facility guidelines. Ensure the proper pressure is set for
the catheter and other disposables connected to the system.
1
(See"Section 15.4.1 - Fluid Delivery Setup Configurable Items"
Figure 10 - 2: Enter Parameters
"Section 10.2 - Save a
Protocol") and/or it can be used for an injection (see
Injecting").
Figure 10 - 3: Selecting Pressure Limit
MEDRAD® MRXperion
for more
2