11.1 Control Room Preparation

11.1.1 Prepare Injection Protocol

A protocol must be set and locked prior to installing syringes in order to use any enabled automatic syringe installation
features.
1.
Set the protocol.
a.
To recall a protocol, see
b. To create or edit a protocol, see
2.
Select Lock on the display.
NOTE: The icon on the display changes to Arm, and the injector head Protocol Lock indicator illuminates.

11.2 Scan Room Preparation

The disposables tray is designed to fit in the handle of the injector for your convenience. If desired, place the tray in the
handle as shown below while installing the syringe and connector tubing or to collect priming waste.

11.2.1 Installing a Syringe

NOTICE
Mechanical Hazard - Equipment damage may result.
Do not install syringes and connector tube with excessive force.
Q
1.
Remove the syringes from packaging.
2.
Install a new syringe on the injector head by inserting it quickly and firmly in one motion until it is securely in place
and an audible click is heard. Remove the dust covers from the tip of each syringe.
NOTE: Syringe A (65 mL) is the contrast syringe and Syringe B (115 mL) is the saline syringe.
"Section 10.3 - Recall a Saved
Protocol".
"Section 10.1 - Create or Edit a
Protocol".