Warnings, Cautions, And Notices; Warnings - Bayer HealthCare Medrad MRXperion Operating Manual

Mr injection system
Hide thumbs Also See for Medrad MRXperion:
Table of Contents

Advertisement

Operation Manual
3

Warnings, Cautions, and Notices

3.1 Warnings

WARNINGS
Air Embolism Hazard - Serious patient injury or death may result.
Use only Bayer approved syringes, connector tubing, and spikes.
Q
Use only catheters and connectors with pressure ratings that are compatible with this system.
Q
Use only accessories and options approved by Bayer.
Q
The operator needs to check for air to confirm that the syringes and tubing have been inspected for the presence of air.
Q
Environmental Contamination Hazard - Serious patient or worker injury or death may result.
Visually inspect all components before use.
Q
Do not use damaged components.
Q
Electric Shock Hazard - Serious patient and/or worker injury or death may result.
Only use the power cord supplied with the system or by the Bayer Services representative.
Q
Equipment must only be connected to a supply mains with protective earth.
Q
Do not remove or open any enclosure, as hazardous voltages exist within system components.
Q
Do not immerse any system components in water or soap solutions.
Q
Disconnect the system from supply mains before cleaning.
Q
Do not use the system if any worn or damaged cabling is detected. Contact Bayer HealthCare Services for replacement.
Q
Fire Hazard - Serious patient and/or worker injury or death may result.
Use the correct fuse type.
Q
Fuse should only be replaced by Bayer or Bayer-trained personnel.
Q
Explosion Hazard - Serious patient and/or worker injury or death may result.
The system contains a lithium ion battery pack. Replacement and disposal shall be performed only by a qualified service engineer.
Q
Please contact Bayer HealthCare Services for assistance.
Compromised Efficacy Hazard - Serious patient injury or death may result.
Do not use this system to deliver any fluid other than intravenous MR contrast media and saline.
Q
Do not retract pistons with connector tube installed.
Q
Procedure Delay Hazard - Serious patient and/or worker injury or death may result.
Turn off any other equipment that could generate high electromagnetic fields or high-level electrostatic discharge.
Q
Personal injury or equipment damage may result when standing on the Power Supply Enclosure.
Do not stand on the power supply enclosure.
Q
3 - 7