System Overview

4.1 System Diagram

The system is comprised of a scan room unit (injector) and a control room unit (touch screen display and Pod). These
components are connected by a fiber optic communications link.
NOTE: When using the system with an open-bore scanner, it is recommended that the injector be placed a
minimum of 60 inches (152.4 cm) from the facade of the scanner.
NOTE: When using the system with a Siemens MAGNETOM 1.5T or 3.0T scanner, it is recommended that the
injector be placed a minimum of 18 inches (45.72 cm) from the facade of the scanner.
NOTE: Consider limits on the length of the connector tubing and patient position when placing the injector.
NOTE: Do not place the scan room unit in a magnetic field greater than 1000G without the casters being locked.
Figure 4 - 1: System Diagram - Power Supply in Scanner Room
1 - Scan Room Unit (Injector)
3 - Scan Room Unit Power Supply
4 - Fiber Optic Communication Link
5- Fiber Optic Quick Disconnect
Scan Room
2 - Power Cable
Control Room
6 - Control Room Unit (Display and Pod)
7 - Display and Pod Power Supplies
8 - Fiber Optic Communication Link
4 - 9