press Exit
less than 4mS, use digital input DI1.
In pulse mode, when the pump receives a pulse stream, the calculated flow output value
will be displayed. The Start / Stop button must be pressed or Remote Start / Stop
activated for the pump to dose at the calculated flow rate. The Pump Status Indicator
will be illuminated and the pump will operate at the flow rate derived from the pulse
mode settings. At least two pulses must be sensed or the Pulse Not Synced
will appear on the notification bar. If the pulse stream results in a calculated flow rate
greater than the pump's calibrated maximum volume the pump will dose at the maximum
flow rate and the Max Pump Volume Exceeded
bar until the pulse rate decreases.
4.3 Batch Mode
In batch mode, upon receiving a single pulse, the pump delivers the programmed
volume of fluid in the programmed amount of time.
FIELD
Batch Volume
Units
Dosing Time (Seconds)
Calculated Flow Rate
Pulse Width (Milliseconds)
Volume Remaining
Figure 5: Batch Mode Home Screen
1. From the Batch Mode home screen (Figure 5) press the Up or Down buttons
to go to the Batch Adjust screen (Figure 6).
to exit without modifying the parameters. For devices with pulse widths
ALLOWABLE VALUES
0.01 to 10,000 (current version only 9999.90)
Fl oz/GAL or mL/L depending on units setting
1 to 86400
0.001 to Maximum Capacity
4 to 60
icon will appear in the notification
9
OPERATION
icon
Time Remaining
% Batch Remaining