Selecting A Preset; Changing A Preset - Fisher Scientific Isotemp User Manual

Shaking and general purpose water baths
Table of Contents

Advertisement

Chapter 4
|
Operation
4-6
|
Isotemp

Selecting a Preset

Three presets are available. Press the soft key below
Note RPM appears on Shaker Water Baths only.
T1
37.0°C
T2 50.0°C
T3 55.0°C

Changing a Preset

Note Changing a preset does not automatically select it. Once a change is made you will have to
return to the Preset screen and follow the Selecting a Preset procedure.
Press the soft key below
Water Baths only.
T1
37.0°C
T2 50.0°C
T3 55.0°C
T1:35.0°C
CAL
RTA1 0.0°C
30 RPM
x
Changing the Real Time Adjustment (CAL RTA) requires a calibrated reference thermometer. If the
displayed temperature does not accurately reflect the actual temperature in the reservoir an RTA
value is required.
As an example, if the temperature is stabilized and displaying 20°C but a calibrated reference
thermometer reads 20.5°C, set the RTA to -0.5°C. After you enter an RTA value allow the display
to stabilize before verifying the bath temperature.
Note The RTA applies only to the selected preset.
To select the desired preset use the arrow keys to
30 RPM
highlight it and then press
50 RPM
will appear with the desired preset set point value.
70 RPM
to display the Presets screen. Note RPM appears on Shaker
To change a preset first highlight it using the arrow keys
30 RPM
and then press
50 RPM
70 RPM
Press the arrow keys to highlight the desired preset
parameter and press
and hold, the arrow keys to change the value.
Once the desired value is displayed press
If desired, press the arrow keys to highlight another
preset parameter or press
home screen.
to display the Presets screen.
. The home screen
.
again. Then press,
to return to the