Download Print this page

Recovering Data - Teledyne WorkHorse Deployment Manual

Monitor / sentinel
Hide thumbs Also See for WorkHorse:

Advertisement

June 2018

Recovering Data

R
D
E C O V E R I N G
A T A I N C L U D E S T H E F O L L O W I N G S T E P S
Downloading data files
Viewing data using the Velocity software
Downloading Data Files
To download a data file:
1. Setup the communication parameters between TRDI Toolz and the ADCP.
2. Wake up the ADCP by pressing the
3. Click Tools, Recover Recorder.
4. TRDI Toolz sends the "r?" and "m?" commands to verify the recorder type and then uses the RY or
MY command to recover the data. The RY and MY command uploads the entire contents of the
recorder via the serial interface to a host computer using the standard YMODEM protocol for bi-
nary file transfer. The data is transferred to the host and stored as binary files.
5. Once the data has been recovered, the recorder can be erased by sending the command RE ErAsE.
This command is case sensitive. Once erased, the data is not recoverable.
Opening a Data File with Velocity
When a *.PD0 (pd zero, not the letter o) file is opened, Velocity creates a matching *.pdv and *.pjv files
depending on the options selected. The *.pdv file is used to process data and the *.pjv file contains the lat-
est information about user selections for processing parameters and data displays.
The original *.pd0 file is never changed, moved, or overwritten.
To open a data file:
1. Start Velocity.
2. Click the Home button (
3. Click the Options button and select the options. Data averaging is on by default.
4. Do one of the following:
Drag a data file onto the Velocity desktop icon. This will start Velocity and open the data file.
With the Start menu selected, on the Starting actions area, click Open a data file button.
Click the Open menu button.
5. After the file is opened, a mini preview of the data file will display at the top of the Velocity screen.
Click the preview to switch to the session (if other files are opened).
See the Velocity quick reference card. A PDF copy is included when you install the soft-
ware/documentation CD.
Page 14
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
button.
) located in the top left corner.
WorkHorse Monitor/Sentinel Deployment Guide
:

Hide quick links:

Advertisement

loading