Gmp Calibration Printout - OHAUS CKW-55 Instruction Manual

Ckw series scale
Table of Contents

Advertisement

EN-44

4.1.5 GMP Calibration Printout

If any setting in the Print-Content-GMP menu (refer to Section 3.6.4 Print Menu) is turned on, GMP automatically prints the
calibration data after calibration or calibration test is completed. Other GMP data, if turned on, will also be printed as shown.
Span Calibration Printout
When performing a Span calibration with any GMP setting turned on, a printout is
automatically made after the procedure is completed.
Linearity Calibration Printout
When performing a Linearity calibration with any GMP setting turned on, a printout is
automatically made after the procedure is completed.
Calibration Test Printout
When performing a Calibration Test with any GMP setting turned on, a printout is
automatically made after the procedure is completed.
CKW-55 CHECKWEIGHER
- - - - - SPAN CAL - - - - -
Cal:
40 kg
Old:
39.99 kg
Diff:
0.01 kg
Wt. Ref. _____________
12/01/04
1:10:00 PM
USER ID 654321
PROJ ID 111111
SCALE ID 123456
Name _______________
- - - - - END - - - - -
- - - - - LIN CAL - - - - -
Cal:
40 kg
Old:
39.99 kg
Diff:
0.01 kg
Wt. Ref. _____________
12/01/04
1:10:00 PM
USER ID 654321
PROJ ID 111111
SCALE ID 123456
Name _______________
- - - - - END - - - - -
- - - - - CAL TEST - - - - -
Test:
10 kg
Act:
10.01 kg
Diff:
0.01 kg
Wt. Ref. _____________
12/01/04
1:30:00 PM
USER ID 654321
PROJ ID 111111
SCALE ID 123456
Name _______________
- - - - - END - - - - -