OHAUS 3000 Series Instruction Manual
OHAUS 3000 Series Instruction Manual

OHAUS 3000 Series Instruction Manual

Indicators
Hide thumbs Also See for 3000 Series:
Table of Contents

Advertisement

3000 Series Indicators
Instruction Manual
T32ME Indicator
T32MC Indicator
i
1.800.561.8187
information@itm.com
www.
.com

Advertisement

Table of Contents
loading

Summary of Contents for OHAUS 3000 Series

  • Page 1 3000 Series Indicators Instruction Manual T32ME Indicator T32MC Indicator 1.800.561.8187 information@itm.com www. .com...
  • Page 2: Table Of Contents

    3000 Series Indicators EN-1 TABLE OF CONTENTS INTRODUCTION ............................EN-4 1.1 Safety Precautions .............................EN-4 1.2 Overview of Parts and Controls ........................EN-5 1.3 Control Functions ............................EN-9 INSTALLATION ............................EN-10 2.1 Unpacking ..............................EN-10 2.2 External Connections ..........................EN-10 2.2.1 RS232 Interface Cable to the indicator ....................EN-10 2.2.2 AC Power .............................EN-10...
  • Page 3 EN-2 3000 Series Indicators TABLE OF CONTENTS (Cont.) 3.5 Readout Menu ............................EN-21 3.5.1 Reset ............................EN-21 3.5.2 Stable ............................EN-21 3.5.3 Filter ............................EN-21 3.5.4 Auto-Zero Tracking ........................EN-21 3.5.5 Sleep (T32ME) ..........................EN-22 3.5.6 Light (T32ME) ..........................EN-22 3.5.7 Backlight (T32MC) ........................EN-22 3.5.8 Auto Off Timer ..........................EN-22 3.5.9 Expand Mode (For testing purpose only) ..................EN-22...
  • Page 4 3000 Series Indicators EN-3 TABLE OF CONTENTS (Cont.) 3.10 End Menu ..............................EN-28 3.11 Security Switch ............................EN-28 OPERATION .............................EN-28 4.1 Turning Indicator On/Off ..........................EN-28 4.2 Zero Operation ............................EN-28 4.3 Manual Tare ............................EN-28 4.4 Changing Units of Measure ........................EN-29 4.5 Printing Data ............................EN-29 4.6 Application Modes ...........................EN-29...
  • Page 5: Introduction

    EN-4 3000 Series Indicators INTRODUCTION This manual contains installation, operation and maintenance instructions for the T32M Indicators. Please read this manual completely before installation and operation. Safety Precautions 1.800.561.8187 information@itm.com www. .com...
  • Page 6: Overview Of Parts And Controls

    3000 Series Indicators EN-5 Overview of Parts and Controls TABLE 1-1. T32MC/T32ME PARTS. Item Description Data Label Front Housing Adjusting Knob (2) Control Panel Mounting Bracket Screw (4) Rear Housing Data Label Power receptacle Strain Relief for Load Cell Cable RS232 Connector Figure 1-1.
  • Page 7 EN-6 3000 Series Indicators Overview of Parts and Controls (Cont.) LFTON LFTOFF Figure 1-2. Main PC Board. TABLE 1-2. MAIN PC BOARD. Item Description Line Power Input J1 LFT On / Off Switch RS232 Connector J3 Sense Jumper W2 Load Cell Terminal Block J7 Sense Jumper W1 1.800.561.8187...
  • Page 8 3000 Series Indicators EN-7 Overview of Parts and Controls (Cont.) Figure 1-3. Controls and Indicators T32MC TABLE 1-3. CONTROL PANEL. Designation Capacity Label Window Pound indicator Ounce indicator Kilogram, gram indicator TARE Menu button CALIBRATION MODE indicator FUNCTION Mode button...
  • Page 9 EN-8 3000 Series Indicators Overview of Parts and Controls (Cont.) Figure 1-4. Controls and Indicators T32ME. TABLE 1-4. CONTROL PANEL. Designation Capacity Label Window Gram indicator Kilogram indicator Pound indicator Ounce indicator TARE Menu button CALIBRATION MODE indicator NET indicator...
  • Page 10: Control Functions

    3000 Series Indicators EN-9 Control Functions TABLE 1-5. CONTROL FUNCTIONS. Button Primary Function ON/ZERO PRINT FUNCTION TARE (Short Press) If Indicator is On, sets Sends the current value Initiates an application Performs a tare zero. to the COM port if mode.
  • Page 11: Installation

    EN-10 3000 Series Indicators INSTALLATION Unpacking External Connections 2.2.1 RS232 interface Cable to the indicator Connect the optional RS232 cable to the RS232 connector Figure 1-1, item 11). Connection Figure 2-1. RS232 Pins. 2.2.2 AC Power Connect the AC Adapter to the power receptacle (Figure 1-1, item 9), then plug the AC Adapter into an electrical outlet.
  • Page 12: Internal Connections

    3000 Series Indicators EN-11 Internal Connections 2.3.1 Opening the Housing CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN ELECTRICAL TECHNICIAN.
  • Page 13: Mounting Bracket

    EN-12 3000 Series Indicators 2.4 Mounting Bracket Attach the bracket to a wall or table using fasteners (not supplied) that are appropriate for the type of mounting surface. The bracket will accommodate up to 6 mm (1/4”) diameter screws. Locate the mounting holes as shown in Figure 2-3.
  • Page 14: Settings

    3000 Series Indicators EN-13 SETTINGS Menu Structure TABLE 3-1. MENU STRUCTURE. PRINT MENU LOCK CALIBRATION SETUP READOUT MODE UNIT SPAN RESET RESET RESET RESET RESET RESET LINEARITY STABILITY 0…31 LEGAL FOR TRADE COUNT KILOGRAM BAUD LOCK CAL 0.5d END CAL 300, …19200...
  • Page 15: Menu Navigation

    EN-14 3000 Series Indicators Menu Navigation TO ENTER THE MENU MODE Press and hold the Menu button until MENU appears on the display. The first upper level menu appears on the display. Allows entry into the displayed menu. Accepts the displayed setting and advances to the next menu item.
  • Page 16: Span Calibration

    3000 Series Indicators EN-15 3.3.1 Span Calibration Span Calibration uses two points to adjust the scale. The first point is the zero value where there is no weight on the scale. The second point is the Span value where a calibration mass is placed on the scale.
  • Page 17: Geographical Adjustment Factor

    EN-16 3000 Series Indicators 3.3.3 Geographical Adjustment Factor The Geographcial Adjustment Factor (GEO) is used to compensate for variations in gravity. Settings from 0 to 31 are available. Refer to table 3-2 to determine the GEO factor that corresponds to your location.
  • Page 18 3000 Series Indicators EN-17 TABLE 3-2. GEOGRAPHICAL ADJUSTMENT VALUES Elevation in meters 1300 1625 1950 2275 2600 2925 3250 1300 1625 1950 2275 2600 2925 3250 3575 Elevation in feet 1060 2130 3200 4260 5330 6400 7460 8530 9600 10660...
  • Page 19: Setup Menu

    EN-18 3000 Series Indicators Setup Menu When the Indicator is used for the first time, enter this menu to set the Reset No, Yes Capacity and Graduation. Legal for Trade Off, On Calibration Unit kg, lb Capacity 5…20000 Graduation 0.0005…20...
  • Page 20 3000 Series Indicators EN-19 TABLE 3-3. SETUP AND CALIBRATION VALUES Capacity Graduation size Graduation size with LFT ON Linearity calibration points with LFT OFF 0.0005, 0.001, 0.002, 0.005 0.001, 0.002, 0.005 2, 5 0.0005, 0.001, 0.002, 0.005, 0.01 0.002, 0.005, 0.01 5, 10 0.001, 0.002, 0.005, 0.01...
  • Page 21: Graduation

    EN-20 3000 Series Indicators 3.4.5 Graduation Set the scale readability. 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20. NOTE 3.4.6 Power On Unit Set the unit of measure displayed at startup. Auto = last unit in use when the indicator was turned off.
  • Page 22: Readout Menu

    3000 Series Indicators EN-21 Readout Menu Enter this menu to customize display functionality. Reset No, Yes Stable 0.5d, 1d, 2d, 3d Filter Level Lo, Med, Hi Auto-Zero Tracking Off, 0.5d, 1d, 3d Sleep (T32ME) Off, On Light (T32ME) Lo, Med, Hi...
  • Page 23: Sleep (T32Me

    EN-22 3000 Series Indicators 3.5.5 Sleep (T32ME) Set the Sleep functionality.(display powers off after 60 seconds of inactivity) = Disabled. = Enabled. 3.5.6 Light (T32ME) Set the brightness of LED display. = Low Intensity = Medium Intensity = High Intensity 3.5.7 Backlight (T32MC)
  • Page 24: Mode Menu

    3000 Series Indicators EN-23 Mode Menu Enter this menu to activate the desired application modes. Reset No, Yes Count Off, On Hold Off, On Check Off, On Totalize Off, On End Mode Exit MODE menu 3.6.1 Reset Set the Mode menu to the factory defaults.
  • Page 25: Unit Menu

    EN-24 3000 Series Indicators Unit Menu Enter this menu to activate the desired units. Default settings are bold. Reset No, Yes Kilogram Off, On Pound Off, On Gram Off, On Ounce Off, On Pound Ounce Off, On End Unit Exit UNIT menu 3.7.1 Reset...
  • Page 26: End Unit

    3000 Series Indicators EN-25 3.7.7 End Unit Advance to the next menu. Print Menu Enter this menu to define printing parameters. Default settings are bold. Reset No, Yes Baud Rate 300, 600, 1200, 2400, 4800, 9600, 19200 Parity 7 Even, 7 Odd, 7 None, 8 None...
  • Page 27: Print Stable Data Only

    EN-26 3000 Series Indicators 3.8.4 Stop Bit Set the number of stop bits. = 1 stop bit. = 2 stop bits. 3.8.5 Handshake Set the flow control method. = no handshaking. ON-OFF = XON/XOFF software handshaking. 3.8.6 Print Stable Data Only Set the print critera.
  • Page 28: Lock Menu

    3000 Series Indicators EN-27 Lock Menu Use this menu to prevent unauthorized changes to menu settings. Default settings are bold. Reset No, Yes Lock Calibration Menu Off, On Lock Setup Menu Off, On Lock Readout Menu Off, On Lock Mode Menu...
  • Page 29: Lock Print

    EN-28 3000 Series Indicators 3.9.7 Lock Print Set the status. = Print menu is not locked. = Print menu is locked and hidden 3.9.8 End Lock Advance to the next menu. 3.10 End Menu Exit to weighing. 3.11 Security Switch A security switch is located on the Main PCB board.
  • Page 30: Changing Units Of Measure

    3000 Series Indicators EN-29 Changing Units of Measure Press and hold the PRINT Units button until the desired measuring unit appears. Only measuring units enabled in the Unit Menu will be displayed (refer to Section 3.7). Printing Data Menu are set (refer to Section 3.8).
  • Page 31 EN-30 3000 Series Indicators Recalling a Stored APW Press the No button to recall the existing APW, then proceed to counting operation. Establishing the Average Piece Weight (APW) The display shows Put 5 Pcs. Establishing a New APW Press the No button to increment the sample size. Choices are 5, 10, 20, 50, 100 and 200.
  • Page 32: Dynamic Weighing

    3000 Series Indicators EN-31 4.6.3 Dynamic Weighing Use this mode to weigh moving or oversized objects. The weight is held on the display until reset. Manual method is available (refer to Section 3.6.3). Begin Dynamic Weighing Press the FUNCTION Mode button to start measurement.
  • Page 33 EN-32 3000 Series Indicators Editing the Over Setting The display shows SEt.HI. Press the Yes button to edit the Over setting. -99999 to 999999 Refer to Menu Navigation Section 3.2 to enter settings. Begin Check Weighing The appropriate Under, Accept or Over LED lights to indicate Check Weigh status.
  • Page 34: Totalize Weighing

    3000 Series Indicators EN-33 4.6.5 Totalize Weighing Use this mode to store the total of a series of weight measurements. NOTE Accumulation (ACC) When the Mode button is released, CLr.Acc Accum is displayed. Start Totalize Press the Yes button to clear the stored data and start new totalize.
  • Page 35: Serial Communication

    EN-34 3000 Series Indicators SERIAL COMMUNICATION The setup of RS232 operating parameters are fully explained in Section 3.8. The physical hardware connection is explained in in Section 2.2. The interface enables display data to be sent to a computer or printer. A computer can be used to control some functions of the indicator using the commands listed in Table 5-1.
  • Page 36: Output Format

    3000 Series Indicators EN-35 Output Format The default serial output format is shown below. Field: Polarity Space Weight Space Unit Stability Legend Length: Weight, up to 6 numbers and 1 decimal, right justified, leading zero blanking. Units, up to 5 characters.
  • Page 37: Legal For Trade

    EN-36 3000 Series Indicators LEGAL FOR TRADE When the indicator is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with met. Settings 1. Verify that the menu settings meet the local weights and measures regulations.
  • Page 38: Sealing

    3000 Series Indicators EN-37 Sealing 6.3.1 Physical Seals For jurisdictions that use the physical sealing method, the local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings. Refer to the illustrations below for sealing methods.
  • Page 39: Audit Trail Seal

    EN-38 3000 Series Indicators 6.3.2 Audit Trail Seal For jurisdictions that use the audit trail sealing method, the local weights and measures official or authorized service agent must record the current configuration and calibration event counter values at the time of sealing. These values will be compared to values found during a future inspection.
  • Page 40: Maintenance

    3000 Series Indicators EN-39 MAINTENANCE CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING. Indicator Cleaning Troubleshooting TABLE 7-1. TROUBLESHOOTING. SYMPTOM PROBABLE CAUSE(s) REMEDY Unit will not turn on. Power cord not plugged in or properly Check power cord connections. Make sure connected.
  • Page 41: Service Information

    If the troubleshooting section does not resolve your problem, contact an authorized Ohaus Service Agent. For Service assistance Service Specialist will be available to assist you. Outside the USA, please visit our website www.ohaus.com to locate the Ohaus office nearest you.
  • Page 42: Technical Data

    3000 Series Indicators EN-41 TECHNICAL DATA Specifications Materials Ambient conditions at 40°C. TABLE 8-1. SPECIFICATIONS Indicator T32MC T32ME Capacity Range 5 to 20000 kg or lb Maximum Displayed Resolution Maximum Approved Resolution Minimum Average Piece wt.(APW) Weighing Units kg, lb, g, oz, lb-oz...
  • Page 43: Accessories

    EN-42 3000 Series Indicators Accessories TABLE 8-2. ACCESSORIES. DESCRIPTION PART NUMBER Interface Cable/PC 25-pin, T32M 80500524 Interface Cable/PC 9-pin, T32M 80500525 Drawings and Dimensions Figure 8-1. T32M Indicator Overall Dimensions. 1.800.561.8187 information@itm.com www. .com...
  • Page 44: Compliance

    3000 Series Indicators EN-43 Compliance Compliance to the following standards is indicated by the corresponding mark on the product. Marking Standard This product conforms to the EMC directive 2004/108/EC, the Low Voltage Directive 2006/95/EC and the Non-automatic Weighing Instruments Directive 2009/23/EC. The complete Declaration of Conformity is available online at www.ohaus.com.
  • Page 45 Industry Canada Note This Class A digital apparatus complies with Canadian ICES-003. ISO 9001 Registration In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International 1.800.561.8187 information@itm.com www.
  • Page 46 Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer.

This manual is also suitable for:

T32meT32mc

Table of Contents