3.6.2 Readout menu

The Readout menu is used to configure display parameters and adopt the scale to
environmental conditions.
Reset
This menu item is used to reset the Readout menu to factory defaults. Selections are: No, Yes.
Units
This menu item is used to define which units will be enabled for use. Selections are: kg, lb, g,
oz, lb:oz, percent. In the event that all units are set to Off, the unit selected for calibration will
forced ON. After pressing the Yes button, each unit may be set to OFF or ON.
Note: Percent unit is only available with a checkweigh target value. (See Section 3.6.3 Mode-
Set Point Input and Section 3.5.2 Checkweigh Operation). The displayed value is the percent
relative to the target weight value (example, if a target value is 100 kg, a 100 kg weight is
displayed as 100%, 75 kg is displayed as 75%, 150 kg is displayed as 150%).
Stable Range
This menu item is used to specify a tolerance limit to indicate a Stable condition (Stable
indicator turns on). Selections are: 0.5, 1, 2, 5 (scale divisions).
Averaging Level
This menu item is used to specify how much filtering is used to compensate for vibration or
excessive air currents. Selections are: Lo, Med, Hi.
Auto-Zero Tracking
This menu item is used to define the Auto-Zero Tracking threshold. AZT minimizes the effects of
temperature changes and small disturbances on the zero reading. The scale maintains the zero
display until the threshold is exceeded. Selections are: OFF, 0.5,1, 3 (scale divisions).
Weight Display Brightness Level
This menu item is used to define the brightness of the weight display. Selections are: OFF
(display blank), Lo, Med, Hi.
Bar Graph Brightness Level
This menu item is used to define the brightness of the Bar Graph display. Selections are: OFF,
Lo, Med, Hi.