3.5.3 Library Operation

The CKW-55 Indicator allows the storage of checkweighing data in 20 Library locations, designated ID 01 to ID 20. Each
Library location includes: a Part Number or name, Pre-set Tare, Maximum, Minimum, and Target value. Data entered into the
libraries can be recalled for use, edited or deleted.
The LIBRARY Save button can be locked to prevent unauthorized changes to the data (refer to Section 3.6.8 Lockout Menu).

Summary of Library Key Functions

Use the following buttons to establish checkweigh set point values (Checkweigh Mode and Library is on).
LIBRARY Save
Short Press
1. Initiates the Library sequence, and
2. Accepts the displayed values and advances to the next sequence.
Long Press during part number display or while END is displayed loads the library data, and exits to
active weigh mode.
NEXT +/-
1. During a Library ID display, short press advances to the next higher Library ID.
2. Long press will begin scrolling through Lib ID's more quickly.
3. During Part Number entry, converts a 3 digit ASCII code to an alpha-numeric character.
4. During numeric value entry sequence, short press toggles the displayed value between negative and
positive.
C
1. Short press in numeric entry sequence, clears one digit at a time. A long Press during data entry
erases all digits.
2. While END is displayed, a long press will erase the data in the selected library.
Exit
Exits any Library function.

Entering and Editing Library Data

In the Mode menu, Library should be On. Library data is viewed in the Set Point Input setting as it was entered. In this
example, Set Point Input setting is Variance.
Note: In Library entry mode, LIBRARY Save is used to confirm and advance to the next setting instead of TARGET button. . . . .
Press the LIBRARY Save button to display the last active Library ID.
To choose another Library ID location, press the NEXT +/- button. The next Library ID location
flashes, then the corresponding Part Number.
Display shows a flashing - - - - - - to indicate that there is no existing Part Number specified
for this Library ID location. For improved identification, an identifying part number or product
ID may be entered.
Enter the desired Part Number (up to 6 digits + 6 decimal points) (example 30123.0) using