Circuit Connection
Diagrams
Connecting INOmax DS
Bunnell Life Pulse Circuit
1
1. From Patient Box
2. Cut Green tube at midpoint
(approximately 6 in. from the Life Port Adapter)
Connecting INOmax DS
Bunnell Life Pulse Circuit
1
2
1. Gas Out Tube from Vent
2. 15M X 4.5mm I.D. Adapter
3. 22M/15F X 22M/15F Adapter
4. Injector Module
WARNING:
• The INOmax DS
used with the Bunnell Life Pulse as ventilator flow rates are
normally below the recommended ventilator flows.
• Place the Life Pulse in Standby prior to suctioning the patient to
avoid NO delivery transiently exceeding the set dose by up to
30 ppm. Press ENTER to reestablish ventilation as soon as the
catheter is removed from the airway. This will limit the extent of
over delivery above the NO set dose.
Part No. 20728 Rev-01
2012-09
2
3
4
3
backup mode (250 mL/min.) should not be
IR
30
Sample Tee to the
IR
3. Insert Sample Tee
4. Life Port Adapter
Injector Module to the
IR
5
6
5. 15M X 4.5mm I.D. Adapter
6. 3cm Piece of Green Gas Out Tube
7. One-Way Valve
8. Green Gas Out Tube to Humidifier
4
8
7