Bagging Systems While Using The Injector Module - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

Circuit Connection
Diagrams
Connection to Bagging Systems While Using
the Injector Module
WARNING: To minimize the delivered concentration of NO2,
Part No. 20728 Rev-01
2012-09
the following steps should be taken for use with
the manual resuscitator bags:
• Use the smallest bag adequate to deliver the
desired tidal volume.
• Inspiratory tubing lengths greater than 72
inches should not be used.
• Use the highest fresh gas flow rated (up to 15
L/min) that is practical.
• Use the lowest practical inspired oxygen
concentration.
• After starting fresh gas flow, squeeze the bag
several times to empty residual gas in the
bag prior to using the system to ventilate a
patient.
24