Pre-Use Checkout
E. Perform Backup INOMAX Delivery Test
1. Turn backup delivery ON.
2. Verify "Backup ON"
alarm activates.
3. Verify values.
NO = 14-26 ppm
NO
= <1.0 ppm
2
4. Turn backup delivery OFF.
F. Complete INOmax DS
Set Dose
Acceptable NO
Value
Acceptable NO
FiO
2
Part No. 20728 Rev-01
2012-09
40 ppm
35-45
ppm
<1.5 ppm
Value
2
95% ±3%
Performance Test
IR
1. Ensure oxygen flow of 10 L/min.
2. Set INOMAX dose to 40 ppm
and allow values to stabilize.
3. Verify values (see table).
4. Turn the INOMAX dose to zero.
10