Mallinckrodt INOmax DSIR Pocket Manual page 9

Delivery system
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

1
2
1. O
Flowmeter (Connected to wall/tank)
2
2. O
Tubing
2
3. 15M x 4.5mm Adapter
4. 22M / 15F x 22M / 15F Adapter
5. Injector Module
(Assemble connectors and tubing as shown)
D. Perform Purge and Alarm Verification
1. Ensure cylinder valve is closed.
2. Set O
flowmeter to 10 L/min.
2
3. Purge INOmax DS
a. Set INOMAX dose to 40 ppm
b. "Cylinder Valve Closed" alarm will occur
c. Cylinder gauge pressure should drop to 0 psig
d. Purge is complete when "Low NO/N
4. Open cylinder valve.
5. Turn INOMAX dose to zero (note: The "Set Dose is Zero, Close
Cylinder Valve" indicator will appear).
10
3
5
4
.
IR
Pre-Use Checkout
9
6
6. 300mm of 22mm Hose
7. Gas Sample Tee
8. Patient Gas Sample Line
with Nafion
9. NO/N
Injector Tube
2
10. Injector Module Electrical Cable
Pressure" alarm activates
2
Part No. 20728 Rev-01
9
8
7
2012-09