Oxygen Dilution Chart

For delivery with 800 ppm cylinder of
INOMAX (nitric oxide) for inhalation.
(Illustrative Only)
10
20
40
80
Caution FiO
Please Note:
The calculations on this chart have been determined based on an
800 ppm cylinder of INOMAX (nitric oxide) for inhalation.
This chart is representative of a range of doses available on the
INOmax DS
IR
recommended therapeutic dose.
Calculations are considered estimates and may vary under clinical
conditions.
All numbers have been rounded to the nearest hundredth.
Part No. 20728 Rev-01
2012-09
.21
0.21
0.20
!
0.20
!
0.19
!
less than 21%
2
and doses higher than 20 ppm are not intended as the
Set FiO
.40
.60
0.40
0.59
0.39
0.59
0.38
0.57
0.36
0.54
Actual FiO
18
2
.80
1.00
0.79
0.99
0.78
0.98
0.76
0.95
0.72
0.90