Table 6-7 U1213A Valid Adjustment Reference Input Values - Keysight U1211A User's And Service Manual

Clamp meters
Hide thumbs Also See for U1211A:
Table of Contents

Advertisement

6
Performance Tests and Calibration
Table 6-7
U1213A valid adjustment reference input values (continued)
Function
Range
DC current
Open
40 A
400 A
1000 A
AC current
40 A
400 A
1000 A
Resistance
Short
10 MΩ
400 kΩ
40 kΩ
4 kΩ
400 Ω
Capacitance
Open
4 µF
40 µF
400 µF
4000 µF
Diode
Short
104
Reference input value
OPEN
30 A
300 A
300 A
02.00 A (70 Hz)
30.00 A (70 Hz)
030.0 A (70 Hz)
300.0 A (70 Hz)
50 A (70 Hz)
300 A (70 Hz)
SHORT
OPEN
10.000 MΩ
300.0 kΩ
30.00 kΩ
3.000 kΩ
300.0 Ω
OPEN
0.300 µF
3.000 µF
30.00 µF
300.0 µF
3000 µF
SHORT
Keysight U1211A, U1212A, and U1213A User's and Service Guide
Valid range for reference input
Keep jaw closed without conductor
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Short Ω and COM terminals
Open terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Open terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0 Ω