Emerson Micro Motion Installation Manual page 23

Specific gravity meters / gas specific gravity measurement
Hide thumbs Also See for Micro Motion:
Table of Contents

Advertisement

3.2.4
Wire the intrinsically safe Time Period Signal (TPS) or
Discrete output version using galvanic isolators
Micro Motion provides a galvanic isolator installation kit specific to wiring the Time Period
Signal (TPS) and Discrete versions of the meter in a hazardous area. Contact your local
sales representative or Micro Motion Customer Support at
more information on ordering an isolator kit for your meter.
CAUTION!
The galvanic isolator kit (TPS/Discrete version) provides isolators for connecting the
following outputs. Use the provided isolators with the designated output.
Note
The RS-485 barrier is not isolated.
Output(s)
4–20 mA + HART
Time Period Signal (TPS)
Discrete
Modbus/RS-485
Power
Procedure
1.
Installation Manual
Meter installation and wiring should be performed by suitably trained personnel only in
accordance with the applicable code of practice.
Refer to the hazardous area approvals documentation shipped with your meter. Safety
instructions are available on the Micro Motion Product Documentation DVD and
accessible on the Micro Motion website at www.emerson.com.
Wire the isolators to the appropriate output terminal and pins (see