3. You are then prompted to enter into the 5500A the value read on the measurement
instrument.
Intermixed with these "output and measure" procedures are internal 5500A
calibration procedures that require no action by the operator.
3-6.
DC Volts
Measure the 5500A output using a precision DMM, and enter into the 5500A each of the
measured values listed in Table 3-2 when prompted to do so.
Step
1
2
3
4
5
6
7
3-7.
AC Volts
Measure the 5500A output using a precision ac voltmeter, and enter into the 5500A each
of the measured values listed in Table 3-3 when prompted to do so.
Step
5500A Output (NORMAL)
1
2
3
4
5
6
7
8
9
10
11
12
13
Note
Table 3-2. DC Volts Calibration Steps
5500A Output (NORMAL)
+3.00 V
-3.00 V
+300 mV
+3.00 V
+30 V
+300 V
+1000 V
Table 3-3. AC Volts Calibration Steps
3.2999 V
0.330 V
3.00 V
3.00 V
30 mV
30 mV
300 mV
30 V
30 V
300 V
300 V
1000 V
1000 V
Calibration and Verification
Frequency
100 Hz
100 Hz
500 kHz
9.99 Hz
100 Hz
500 kHz
100 Hz
100 Hz
100 kHz
100 kHz
20 kHz
100 Hz
7 kHz
3
Calibration
3-5