6-12. Trigger Signal Specifications (Edge Function)

Table 6-9. Trigger Signal Specifications (Edge Function)
Edge Signal
Division
Frequency
Ratio
1 kHz to 10 MHz
off/1

6-13. Trigger Signal Specifications (Square Wave Voltage Function)

Table 6-10. Trigger Signal Specifications (Square Wave Voltage Function)
Edge Signal
Division
Frequency
Ratio
10 Hz to 10 kHz
off/1

6-14. Trigger Signal Specifications

Trigger Signal Type
Field Formats
Polarity
Amplitude into 50Ω (p-p)
Line Marker

6-15. Oscilloscope Input Resistance Measurement Specifications

Table 6-12. Oscilloscope Input Resistance Measurement Specifications
Scope input selected
Measurement Range
Uncertainty

6-16. Oscilloscope Input Capacitance Measurement Specifications

Table 6-13. Oscilloscope Input Capacitance Measurement Specifications
Scope input selected
Measurement Range
Uncertainty
[1] Measurement made within 30 minutes of capacitance zero reference. SC600 option must be
selected for at least five minutes prior to any capacitance measurement, including the zero process.
Typical Amplitude
into 50Ω (p-p)
≥ 1 V
Typical Amplitude
into 50Ω (p-p)
≥ 1 V
Table 6-11. TV Trigger Signal Specifications
Selectable NTSC, SECAM, PAL, PAL-M
Selectable inverted or uninverted video
Adjustable 0 to 1.5 V p-p into 50 ohm load, (±7% accuracy)
Selectable Line Video Marker
50 Ω
40 Ω to 60 Ω
0.1 %
± (5% of input + 0.5 pF) [1]
Typical Rise Time
Typical Lead Time
≤ 2 ns
40 ns
Typical Rise Time
Typical Lead Time
≤ 2 ns
1 µs
Parameters
1 MΩ
500 kΩ to 1.5 MΩ
0.1 %
1 MΩ
5 pF to 50 pF
SC600 Option
6
SC600 Specifications
6-11