General Specifications - Fluke 5500A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Service Manual
1-4.

General Specifications

Warmup Time
Settling Time
Standard Interfaces
Temperature
Performance
Temperature Coefficient
Relative Humidity
Altitude
Safety
Analog Low Isolation
EMC
Line Power
Power Consumption
Dimensions
Weight (without options)
Absolute Uncertainty
Definition
Specification
Confidence Interval
[1] After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be
required.
1-6
Twice the time since last warmed up, to a maximum of 30 minutes.
Less than 5 seconds for all functions and ranges except as noted.
IEEE-488 (GPIB), RS-232, 5725A Amplifier
• Operating: 0 °C to 50 °C
• Calibration (tcal): 15 °C to 35 °C
• Storage: -20 °C to 70 °C
Temperature Coefficient for temperatures outside tcal +5 °C is 0.1X/°C of
the 90-day specification (or 1-year, as applicable) per °C.
• Operating: <80% to 30 °C, <70% to 40 °C, <40% to 50 °C
[1]
• Storage: <95%, non-condensing
• Operating: 3,050 m (10,000 ft) maximum
• Non-operating: 12,200 m (40,000 ft) maximum
Complies with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
20 V
Designed to comply with FCC Rules Part 15; VFG 243/1991. If used in
areas with Electromagnetic fields of 1 to 3 V/m, resistance outputs have a
floor adder of 0.508 Ω. Performance not specified above 3 V/m. This
instrument may be susceptible to electro-static discharge (ESD) from
direct contact to the binding posts. Good static aware practices should be
followed when handling this and other pieces of electronic equipment.
• Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
• Line Frequency: 47 Hz to 63 Hz
• Line Voltage Variation: ±10% about line voltage setting
5500A Calibrator, 300 VA; 5725A Amplifier, 750 VA
5500A Calibrator:
• Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm (0.6
inch) for feet on bottom of unit;
• Width: 43.2 cm (17 inches), standard rack width
• Depth: 47.3 cm (18.6 inches) overall 5725A Amplifier:
• Height, 13.3 cm (5.25 inches), standard rack increment, plus 1.5 cm
(0.6 inch) for feet on bottom of unit;
• Width, 43.2 cm (17 inches), standard rack width
• Depth, 63.0 cm (24.8 inches) overall.
5500A Calibrator, 22 kg (49 lb); 5725A Amplifier 32 kg (70 pounds)
The 5500A specifications include stability, temperature coefficient,
linearity, line and load regulation, and the traceability of the external
standards used for calibration. You do not need to add anything to
determine the total specification of the 5500A for the temperature range
indicated.