Step
5500A Output (NORMAL)
330 ρF
1
499 ρF
2
0.5 ηF
3
1.00 ηF
4
1.10 ηF
5
3.2 ηF
6
3.5 ηF
7
10 ηF
8
11 ηF
9
32 ηF
10
35 ηF
11
100 ηF
12
110 ηF
13
320 ηF
14
0.35 µF
15
1 µF
16
1.1 µF
17
3.2 µF
18
3.5 µF
19
10 µF
20
11 µF
21
32 µF
22
35 µF
23
100 µF
24
110 µF
25
320 µF
26
350 µF
27
600 µF
28
Table 3-9. Capacitance Calibration Steps
Calibration and Verification
Recommended Stimulus
2 V rms at 1 kHz
"
"
"
"
"
"
"
"
"
"
"
"
1 V rms at 1 kHz
1 V rms at 100 Hz
"
"
"
"
"
"
"
"
"
"
"
1 V rms at 50 Hz
"
3
Calibration
3-13