Dc Voltage Calibration - Fluke 5500A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

HP 3458A
BNC(F) to
Double Banana
Adapter
Figure 6-19. Equipment Setup for SC300 Square Wave Measurements.
For all measurements, the HP 3458A is in DCV, manual ranging, with level triggering
enabled. A convenient method to make these measurements from the HP 3458A's front
panel is to program these settings into several of the user defined keys on its front panel.
For example, to make topline measurements at 1 kHz, you would set the DMM to
"NPLC .01; LEVEL 1; DELAY .0002; TRIG LEVEL". To find the average of multiple
readings, you can program one of the keys to "MATH OFF; MATH STAT" and then use
the "RMATH MEAN" function to recall the average or mean value. Refer to Figure 6-19
for the proper connections.

6-104. DC Voltage Calibration

This procedure uses the following equipment:
Hewlett-Packard 3458A Digital Multimeter
50 Ω feedthrough termination (as required in the calibration procedure)
Shorted Dual Banana Connector
BNC(f) to Double Banana adapter
BNC cable supplied with the SC300
Full calibration of the Voltage Function requires both dc and ac
calibration.
Refer to Figure 6-19 for the proper setup connections.
Set the Calibrator Mainframe in Scope Cal mode, DC Voltage section. Follow these
steps to calibrate DC Voltage:
1. Connect the Calibrator Mainframe's SCOPE connector to the HP 3458A input, using
the BNC cable and the BNC(f) to Double Banana adapter.
2. Set the HP 3458A to DCV, Auto Range, NPLC = 10, FIXEDZ = on.
3. Press the GO ON blue softkey.
Calibration and Verification of Square Wave Functions
SC300 Cable
50 Ω Feedthrough
Termination
Note
SC300 Option
5500A-SC300
5500A
CALIBRATOR
SCOPE
NORMAL AUX
200V PK
A,
-SENSE,
V,
,
MAX
AUX V
RTD
HI
1000V
20V
RMS
RMS
MAX
MAX
TRIG
LO
OUT
1V PK
MAX
20V PK
MAX
TC
20V PK
MAX
om062f.eps
6
6-79