3-32. AC Voltage Amplitude Accuracy (NORMAL)

The AC Voltage Amplitude Accuracy test verifies the accuracy of ac voltage at the
5500A Calibrator front panel NORMAL terminals. Table 3-18 shows the test points.
Table 3-18. AC Voltage Amplitude Accuracy Test (NORMAL)
Nominal Value
30 mV
9.5 Hz
30 mV
10 Hz
30 mV
45 Hz
30 mV
1 kHz
30 mV
10 kHz
30 mV
20 kHz
30 mV
50 kHz
30 mV
100 kHz
30 mV
450 kHz
300 mV
9.5 Hz
300 mV
10 Hz
300 mV
45 Hz
300 mV
1 kHz
300 mV
10 kHz
300 mV
20 kHz
300 mV
50 kHz
300 mV
100 kHz
300 mV
500 kHz
3 V
9.5 Hz
3 V
10 Hz
3 V
45 Hz
3 V
1 kHz
3 V
10 kHz
3 V
20 kHz
3 V
50 kHz
3 V
100 kHz
3 V
450 kHz
Frequency
Measured Value
(V)
(NORMAL)
Calibration and Verification
Performance Verification Tests
Deviation %
90-Day Spec. (%)
5.550
0.327
0.177
0.177
0.177
0.217
0.257
0.370
0.950
5.550
0.207
0.047
0.047
0.047
0.087
0.133
0.227
0.640
5.550
0.118
0.022
0.022
0.022
0.062
0.110
0.227
0.490
3
3-25