3-48. AC Voltage Harmonic Amplitude Accuracy (AUX)
The AC Voltage Harmonic Amplitude Accuracy (AUX) tests the accuracy of the 50th
harmonic from the AUX terminals. For this test, set the 5500A output to sinewave. Table
3-35 shows the test points.
Table 3-35. AC Voltage Harmonic Amplitude Accuracy (AUX)
Nominal
Nominal
Value
Value
(NORMAL)
(AUX)
100 mV
329 mV
100 mV
329 mV
100 mV
329 mV
100 mV
3.29 V
100 mV
3.29 V
100 mV
3.29 V
3-49. DC Voltage Offset Accuracy
The DC Voltage Offset Accuracy test the accuracy of the dc offset function for an ac
sinewave output on the NORMAL terminals. Table 3-36 shows the test points.
Nominal
Nominal DC
ACV
Value
Value
10 mV
0 V
10 mV
50 mV
100 mV
0 V
100 mV
500 mV
1 V
0 V
1 V
5 V
3.3 V
0 V
3.3 V
45 V
Frequency
Frequency
(AUX)
(NORMAL)
1 kHz
20 Hz
5 kHz
100 Hz
10 kHz
200 Hz
1 kHz
20 Hz
5 kHz
100 Hz
10 kHz
200 Hz
Table 3-36. DC Voltage Offset Accuracy Test
Frequency
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Measured
Deviation
Value (V)
(AUX)
Measured Value
Deviation
(V DC) (NORMAL)
Calibration and Verification
Performance Verification Tests
90-Day Spec
%
(%)
0.305%
0.424
0.574
0.097
0.235
0.385
1-Year Spec.
%
(µV or %)
33 µV
0.166%
330 µV
0.166%
3.3 mV
0.166%
33 mV
0.173%
3
3-39