Section 2: Description & Specifications
September 2013
2.2
Transmitter Performance Specifications
2.2.1
Performance Specifications - Transmitter (pH input)
Measurement Range [pH]: 0 to 14 pH
Accuracy: ±0.01 pH ±1mV @ 25°C ± 0.03 pH
Diagnostics: Glass impedance, Reference impedance, Slope and Offset (mV)
Temperature coefficient: ±0.002pH/ ºC
Solution temperature correction: Pure water, dilute base and custom.
Buffer recognition: NIST, DIN 19266, JIS 8802, and BSI.
Input filter: Time constant 1 - 999 sec, default 4 sec.
Response time: 5 seconds to 95% of final reading
2.2.2
Performance Specifications - Transmitter (ORP input)
Measurement Range [ORP]: -1400 to +1400 mV
Accuracy: ± 1 mV
Temperature coefficient: ±0.12mV/°C
Input filter: Time constant 1 - 999 sec, default 4 sec.
Response time: 5 seconds to 95% of final reading
Recommended Sensors for ORP: All standard ORP sensors
Recommended Sensors for pH: All standard pH sensors. Supports SMART pH sensors from
Rosemount Analytical.
2.3
Basic Fieldbus Specifications
The 1066pH, the first pH transmitter registered to ITK 6, has the following basic features:
DIAGNOSTICS: Full comprehensive sensor, transmitter, and calibration diagnostics, which are
available to the bus via Field Diagnostics.
COMMON SOFTWARE DOWNLOAD SPECIFICATION: Allows software upgrades to be down-
loaded to the transmitter, while it is in service.
LINK ACTIVE SCHEDULER: Allows the 1066pH to function as a Linkmaster.
4 AI FUNCTION BLOCKS: For PV (pH, ORP, Redox), temperature, reference electrode impedance,
and glass electrode impedance.
AO FUNCTION BLOCK: Allows the transmitter to use a temperature measurement from the bus for
temperature compensation.
ADDITIONAL FUNCTION BLOCKS: Are provided for use by the 1066-P-FF or other transmitters on
the segment, and include:
MATH FUNCTION BLOCKS: Arithmetic and Integrator
ANALOG CONTROL FUNCTION BLOCKS: PID Control, Input Selector, Signal Characterizer, Control
Selector, and Output Splitter.
Input: One isolated pH/ORP sensor input.
Power & Load Requirements: Fieldbus power voltage at 9 to 32 VDC. Current required is 18 mA.
4
1066 Transmitter Instruction Manual
Description and Specifications
LIQ_MAN_1066-P-FF