Functions and Features

The following table gives an overview of the important functional and performance characteristics of
the Thermo Scientific Sorvall ST 8 Centrifuge.
Tabelle 1-3. Functions and Features
Component / Function
Structure / Housing
Chamber
Drive
Keys and display
Controls
Internal memory
Functions
Acceleration / braking profiles
Rotor recognition
Imbalance recognition
Lid lock

Mains Supply

The following table contains an overview of the electrical connection data for the Thermo Scientific
Sorvall ST 8 Centrifuge. This data is to be taken into consideration when selecting the mains
connection socket.
Tabelle 1-4. Electrical connection data of the Thermo Scientific Sorvall ST 8 Centrifuge
Cat.
75007221
ventilated
75007220
ventilated
Thermo Scientific
Mains
voltage
Frequency
± 10 %
230 V
50 / 60 Hz
120 V
60 Hz
Description / Features
Galvanized steel chassis with armoured plating
Stainless steel
Induction drive without carbon brushes
Easy-to-clean keypad and display surface
Microprocessor-controlled
The most recent data is saved
RCF-selection
Standard and Soft Acceleration/Deceleration
Automatic / Electric
Electronic, contingent on rotor and speed
Lid is locking, when being pressed down
Rated
Power
current
consumption
2,0 A
310 W
5,0 A
310 W
1
Introduction and Description
Functions and Features
Equipment fuse
Building fuse
5 AT
16 AT
10 AT
15 AT