Functions And Features; Mains Supply - Thermo Scientific Sorvall Legend X1R Instruction Manual

Hide thumbs Also See for Sorvall Legend X1R:
Table of Contents

Advertisement

Functions and features

The following table gives an overview of the important functional and performance characteristics
of the Sorvall Legend X1R.
Table 1-3. Functions and Features
Component / function
Structure / Housing
Chamber
Drive
Keys and display
Controls
Internal memory
Functions
Acceleration / braking profiles
Rotor recognition
Imbalance recognition
Lid lock

Mains supply

The following table contains an overview of the electrical connection data for the Sorvall Legend
X1R. This data is to be taken into consideration when selecting the mains connection socket.
Table 1-4. Electrical connection data of the Sorvall Legend X1R
Mains
Cat.
voltage
230 V ±10 %
refrigerated
1
Please contact the Thermo Fisher service to run the unit at 60Hz Line frequency.
Thermo Scientific
Description / features
Galvanized steel chassis with armoured plating
Stainless steel
Induction drive without carbon brushes
Easy-to-clean keypad and display surface
Microprocessor-controlled
The most recent data is saved
RCF-selection, temperature control, pretemp with cooling equipped
devices
9 acceleration and 10 braking curves
Automatic
Electronic, contingent on rotor and speed
Automatic lid closing and locking starting from an initial hold position
Frequency
Rated current
50 / 60 Hz
8 A
1
Power
Equipment
consumption
fuse
1400 W
15 AT
Sorvall Legend X1R
Introduction and description
Functions and features
Building fuse
16 AT