To Remove the Valve Cover

Maintenance
1
Gather the following:
T- 20 Torx driver
2
Remove the pneumatics front cover. See
Pneumatics Front
Remove the valve thermal enclosure. See
3
Valve Thermal
Enclosure".
4
Completely loosen the two T- 20 Torx captive screws on
the valve cover
(Figure
Valve cover
Figure 9
Valve cover
5
Remove the valve cover.
Reinstallation is the reverse of these steps.
Removing Covers and Components
Cover".
9).
2
"To Remove the
"To Remove the
17