To Remove The Tray Assembly (111-Vial Model) - Agilent Technologies 7697A Maintenance Manual

Headspace sampler
Hide thumbs Also See for 7697A:
Table of Contents

Advertisement

2
Removing Covers and Components

To Remove the Tray Assembly (111-Vial Model)

20
1
Gather the following:
Lint- free gloves
2
Set your GC oven, Headspace oven, and transfer line to
ambient temperatures and wait for them to cool.
Once the GC oven, Headspace oven, and transfer line have
3
cooled to ambient temperature, turn the Headspace off
and unplug the power cord.
Turn all gas flows off at their sources.
4
Remove all sample vials and vial racks.
5
Lift the front of the tray assembly up and disconnect the
6
two cable connections underneath the tray
Figure 12 Tray cable connections
(Figure
12).
Cable in clip
Cable connections
Maintenance