To Replace the Sample Loop Adapters

C A U T I O N
N O T E
42
Normally, only change or replace the sample loop adapters
when changing sample loop size. A full set of adapters
shipped with the Headspace.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
The loop cover is used as an example in this procedure. Steps for the
valve/loop cover are similar.
1
Gather the following:
Lint- free gloves
Sample loop (see
Sample loop adapters, as needed (see
Needle nose pliers
Depending on your hardware configuration, remove the
2
loop cover or the valve/loop cover, then remove the
sample loop. Follow the steps in
Loop".
Table
7)
"To Replace the Sample
Table
2)
Maintenance