To Replace The Sample Tray Seals (12-Vial Model) - Agilent Technologies 7697A Maintenance Manual

Headspace sampler
Hide thumbs Also See for 7697A:
Table of Contents

Advertisement

3
Maintenance

To Replace the Sample Tray Seals (12-Vial Model)

WA R N I N G
26
The sample tray seals normally do not require replacement.
However, they may need to be cleaned or replaced after
cleaning up spills or other maintenance.
Vials may be hot. Before handling, allow time for vials to cool as
needed. Wear heat resistant gloves as needed.
1
Gather the following:
Replacement vial tray seals
2
Lift the rotating top cover.
Rotating top cover
3
Remove all vials from the carousel.
4
If needed, remove the old vial tray seals.
5
If needed, clean the tray assembly. See
Sample Tray (12- Vial
Place the new sample tray seals into the vial slots in the
6
tray.
"To Clean the
Model)".
Maintenance

Advertisement

Table of Contents
loading

Table of Contents