To Attach the Transfer Line to a Cool On-Column Inlet

C A U T I O N
WA R N I N G
Maintenance
Make sure your column dimensions are correct in relation to
the size of your fused silica. Your column id can not be
greater than 530 um. Refer to your cool on- column inlet
documentation for more information.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Gather the following:
1
Lint- free gloves
Capillary column cutter (5181- 8836)
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Cool the GC inlet to a safe handling temperature.
2
Cool the GC oven to room temperature to prevent column
3
damage. Then turn off the carrier gas.
Install the transfer line to the 6 port valve. See
4
a Fused Silica Column into the Transfer
Trim the fused silica so that it extends 42 mm out of the
5
inner metal sleeve on the transfer line.
Slide the fused silica through the retaining nut on the
6
cool on- column inlet.
Finger- tighten the retaining nut. Pull the transfer line
7
gently to be sure the PTFE ferrule has formed a tight seal
with the transfer line. Tighten the retaining nut further, if
necessary.
3
Maintenance
"To Install
Line".