Six Port Valve Errors
Six port valve motor
Six port valve shaft
Agilent 7697A Troubleshooting
The six port valve rotates during the sampling cycle. If the
valve cannot rotate, power cycle the HS. Check for
obstructions in the pneumatics area (such as whether the
insulating cover interferes with the valve drive shaft). If the
error persists, contact Agilent for service.