To Attach The Transfer Line To A Split Splitless Or Multimode Inlet - Agilent Technologies 7697A Maintenance Manual

Headspace sampler
Hide thumbs Also See for 7697A:
Table of Contents

Advertisement

3
Maintenance

To Attach the Transfer Line to a Split Splitless or Multimode Inlet

C A U T I O N
WA R N I N G
50
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
1
Gather the following:
Lint- free gloves
Two 7- mm wrenches
Capillary column cutter (5181- 8836)
Inlet septum, green (5183- 4759)
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Cool the GC inlet to a safe handling temperature.
2
Cool the GC oven to room temperature to prevent column
3
damage. Then turn off the carrier gas.
4
Install the transfer line to the 6 port valve. See
a Fused Silica Column into the Transfer
Slide the one- piece septum nut and strain relief coupling
5
onto the tubing until the inner metal sleeve of the
transfer line bottoms- out on the strain relief.
Slide the septum onto the tubing until the septum is
6
against the bottom of the one- piece septum nut and
strain relief coupling.
7
Trim the tubing 35 mm past the end of the septum.
8
Install the inlet liner (with o- ring) that is appropriate for
your application.
Insert the tip of the transfer line into the inlet.
9
Finger- tighten the strain relief coupling.
10
"To Install
Line".