To Remove the Pneumatics Assembly

Maintenance
Remove the pneumatics assembly
oven components.
Figure 4
Pneumatics assembly without covers
To remove the pneumatics assembly:
Set your GC oven, Headspace oven, and transfer line to
1
ambient temperatures and wait for them to cool.
2
Once the GC oven, Headspace oven, and transfer line have
cooled to ambient temperature, turn the Headspace off
and unplug the power cord.
3
Turn all gas flows off at their sources, and if necessary,
disconnect the gas lines at the pneumatics assembly back
panel.
4
Remove the pneumatics assembly front cover. See
Remove the Pneumatics Front
Removing Covers and Components
(Figure
4) to access the
Cover".
2
Pneumatics
assembly
"To
13