To Remove the Fused Silica Column from the Transfer Line

C A U T I O N
C A U T I O N
WA R N I N G
Maintenance
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
1
Gather the following:
3/16- inch open end wrench (provided in ship kit)
1/4- inch open end wrench
2
Cool down the transfer line and 6 port valve (sample
loop) in the headspace sampler.
Disconnecting the transfer line may interrupt GC carrier gas flow.
Cool the GC column oven and inlet as needed to prevent column
damage.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Cool the GC inlet to a safe handling temperature.
3
If using GC carrier gas control, or if the Headspace
4
provides the carrier gas flow for the GC, also cool the GC
oven to ambient temperature to prevent column damage.
Remove the pneumatics front cover. See
5
Pneumatics Front
Cover".
3
Maintenance
"To Remove the
59