To Clean the Sample Tray Assembly (111-Vial Model)

C A U T I O N
Maintenance
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
The following describes how to clean the 111- vial model
sample tray assembly.
Gather the following.
1
Lint- free gloves
Lint- free cloth
Vacuum cleaner
Clean the vial racks.
2
Remove the vial racks from the tray assembly
a
(Figure
16).
b
Carefully remove all vials from the vial racks and set
them aside.
Remove any broken glass from the vial racks using a
c
vacuum cleaner.
d
Wipe away any spills from the vial racks using a
lint- free cloth.
Figure 16 Removing vial racks from the tray assembly (111-Vial Model)
3
Maintenance
27