Repair and Replacement Procedures
Removing and Replacing the A22 Splitter
Removing and Replacing the A22 Splitter
Tools Required
— T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
— T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
— 5/16 inch open-end torque wrench (torque to 10 in-lb or 1.13 N.m)
— ESD grounding wrist strap
Removal Procedure
Refer to
Be careful not to damage the center pins of the semirigid cables. Some
flexing of the cables is necessary to remove assemblies but do not
over-bend them.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to
3. Position the analyzer bottom side up.
4. Disconnect cable W42 at the A22 splitter connection.
5. Remove cable W43 from between the A22 splitter and the A23 mixer brick.
6. Remove cable W44 from between the A22 splitter and the A24 mixer brick.
7. Remove two screws (item
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests
Keysight N5221A/22A Service Guide
Figure 7-15
for this procedure.
of the analyzer.
Torque all RF cable connections to 10 in-lbs.
that pertain to this removal procedure. Refer to
"Removing the Covers" on page
①
) from the A22 splitter and lift the splitter out
Table 7-2 on page
7-6.
7-60.
7-35