Ee Default Adjustment; If Response Adjustment (Option 090, 093, Or 094 Only) - Keysight Technologies N5221A Service Manual

2-port and 4-port pna microwave network analyzers, 10 mhz to 13.5 / 26.5 ghz
Hide thumbs Also See for N5221A:
Table of Contents

Advertisement

Tests and Adjustments
Adjustments
Figure 3-18
Equipment Setup 3 for the Receiver Adjustment

EE Default Adjustment

This sets the EEPROM data to their default values.
Procedure
1. Press UTILITY
2. On the dialog box, select Initialize rather than Adjust/Verify because an
3. Follow the instructions and prompts as they are displayed.

IF Response Adjustment (Option 090, 093, or 094 Only)

The IF Response adjustment is required when upgrading to Option 090, 093
and or 094. In addition, this adjustment must be performed each time a new
configuration is set up or if cables are changed:
Procedure
1. Press UTILITY
2. Follow the instructions and prompts as they are displayed.
3-46
System
Default Adjustment.
adjustment is typically unnecessary.
System
Response Adjustment.
, then
Service
, then
Adjustments
Service
Adjustments
, then
, then
Keysight N5221A/22A Service Guide
, then click EE
, then click IF

Advertisement