Performance Tests
4·3 0. HP·IB Check (Optional)
Up to this point the 3585A has been checked only as a bench operated instrument. If the in­
strument is to be used with a controller, the HP-IB interface should be checked. The pro­
gram shown in Figure 4-7 will check the HP-IB operation of the instrument to a high level of
confidence. This program is flow charted using controller independent language so that it
may be adapted to your controller. If you have a -hp- 9825A calculator, a listing of this pro­
gram appears in Table 4-3 . the program is also contained on File 26, Track
Automatic Performance Test tape (P.N. 03585-1000 1 ). If an error is detected in the HP-IB
interface of the 3585A, an error number will be printed out. The error definitions are con­
tained in Table 4-2 and may be used to help locate problems on the 3585A HP-IB board.
To run the HP-IB check with the -hp- 9825A calculator, insert the Semi-Automatic Perfor­
mance Test tape in the calculator tape slot and press the following keys:
When the lazy "T"
key. To complete the test, follow the instructions on the calculator display. If no HP-IB
errors are found by the test, "HP-IB OK" will be printed by the calculator. This ends the
HP-IB check program.
4-3 2
( �
) has reappeared on the 9825A display, press the
Table 4·2. Hp·IB Error Definitions
#
Error
1
2
Large HP-IB Problem; DSA Required
3
4
5
Data Line Problem
6
7
8
9
Front Panel Light or I nterface Prob-
1 0
lem; otherwise use DSA
1 1
m
Explanation
Model 3585A
0
of the Semi­
8

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents