Model 3585A
5·36. Flatness Adjustment
a. Move test jumper A 1 5W 1 to the "TEST" position.
b. Set the 3585A controls for:
c. Using the Continuous Entry control, adjust the REF LVL so that the trace is centered
on the CRT.
d. The 3585A is now in its 0.2dB/DIV mode. This allows very fine adjustment of the in
struments flatness.
e. Adjust the input flatness with the following components in the order shown.
The input flatness of the instrument should resemble Figure 5-39 when completely adjusted.
The effect of each adjustment is shown in Figure 5-40 (foldout). Continue adjustment of the
instrument flatness until the peak to peak variation of the trace is less than 0.2dB
division).
f. Move test jumper A 1 5W 1 to the "NORM" position.
REF -25. 0 dBm
10 dB/O I V
RANGE -25. 0 dBm
- - - -
- - - -
- - - -
- - - -
- - - -
CENT E R 20 000 000. 0 H z
RBW 30 KHz
VBW 30 KHz
Figure 5·38. Normal Display For Test Mode 05
RECALL 605
INSTRUMENT PRESET
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . 20. 1 MHz
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2dB
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 25dBm
AUTO RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
A 1 R 1 3 1 , C83, L 1 8, C86, L 1 9, C89, L21 , C92
MARKER 20 000 (JOO. O Hz
-37. 1
dBm
t
,
,
-
- - - -
- - - -
- - - -
. - - - -
-
- -
SPAN 4 0 000 000. 0 Hz
ST . 2 SEC
Adjustments
(l
5-5 9