Installation and Interfacing
OPTION 907
.
I.LASTlC TRI M
H A N D L E K I T , B E NC H O P E R A T I O N
2· 1 3. Rack Mounting Without Slides
a. Remove the plastic trim (Figure 2-2) and front handles from the 3585A if it is so equip­
ped.
b. Remove the plastic feet from the bottom of the 3585A.
c. Install the Rack Flange Kit with or without handles according to the instructions in­
cluded in the kit:
Rack Flange Kit (no handles) . . . . . . . . . . Option 908, -hp- Part Number 5061 -()()79
Rack Flange
d. Install an Instrument Support Rail on each side of the instrument rack. (The Instru­
ment Support Rails, used to support the weight of the instrument, are included with -hp­
rack-mount cabinets.)
1. The weight of the 3585A must be supported by Instrument sup­
port Rails inside the instrument rack. Do not under any cir­
cumstances attempt to rack mount the 3585A using only the front
flanges.
2.
Use extreme care when lifting it to avoid personal injury.
e. Using two people, lift the 3585A to its position in the rack on top of the Instrument
Support Rails.
f. Using the appropriate screws, fasten the 3585A's Rack-Mount Flanges to the front of
the instrument rack.
2-4
.--- --
-
RACK M O U N T F LA N G E K I T
Figure 2·2. Rack Mount and Handle Kits
&
Front Handle Kit . . . . . . . Option 909, -hp- Part Number 5061 -()()85
WARNING
The 3585A is heavy for its size (approximately 88 lbs, 40 kg.).
R A C K M O U NT F LA N G E / F R O N T H A N D L E K I T