Adjustments
d. Replace the HP-IB board back in the card nest.
e. Turn the 3585A power on.
f. Connect a DVM to A44TP5. Set the DVM for the 20Vdc range.
g. Adjust A44R9 (see Figure 5-46) for the voltage stamped on A44U 16 ± 0.2V.
h. Disconnect the DVM. This completes the HP-IB Adjustments.

5-48. X-Y Plotter Adjustment

a. Connect a DVM to A62TP l (REF). Set the DVM to the 20Vdc range.
b. Adjust A62R4 (see Figure 5-46) for a dc voltage reading of - 1O.24Vdc ± 0.02V.
c. Disconnect the DVM. This completes the X-Y Plotter Adjustments.
5-70
Model 3585A